A:96
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
CAUTION!
It is important to set a correct ultrafiltration rate for the patient, to avoid incorrect fluid
removal. Set a minimum ultrafiltration rate according to the instructions for the
specific dialyzer.
CAUTION!
Once treatment parameters are set, you must verify that all calculated treatment
values are as intended.
CAUTION!
Additional measures to supervise the patient weight loss is recommended when
treating low weight patients or when performing long treatments. For Ultrafiltration
details, refer to Technical data and specifications.
Procedure
1)
Press the
Time
indicator.
2)
Press
UF volume
.
3)
Calculate the patient’s ultrafiltration
volume, see Section 4.2.10
4)
Enter the patient’s ultrafiltration volume.
5)
Press
OK
.
4.2.11
Set heparin values
Before you begin
WARNING!
Make sure to set the prescribed heparin values. Incorrect heparin settings may
cause blood coagulation or internal bleeding.
Procedure
1)
Press the
Blood
button.
2)
Select the
Heparin
tab.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...