AK 98 Dialysis Machine - Attentions
B:95
520 Concentrate tube A is out of position
520 Concentrate tube A is out of position
Check the position.
Appears:
When the red concentrate connector is not in the correct
position.
Machine actions:
After 2 minutes there is a buzzer sound.
Operator action:
Check the placement of the red concentrate connector.
Mute time:
5 minutes.