AK 98 Dialysis Machine - Attentions
B:93
518 CleanCart is not attached
518 CleanCart is not attached
Attach CleanCart cartridge.
Appears:
When the CleanCart cartridge is not properly attached to
the BiCart cartridge holder during a heat disinfection
program.
Machine actions:
After 1 minute there is a buzzer sound. The delay is only
valid at first occurrence.
Operator action:
1.
Attach the CleanCart cartridge to the BiCart cartridge
holder.
2.
If it is already attached, make sure it is properly located.
“Start a heat disinfection with a
handbook.
Mute time:
Permanent.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...