B:90
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
515 Check rotor/tube distance
515 Check rotor/tube distance
To continue press Confirm.
Appears:
When the blood pump segment has been changed.
Machine actions:
None.
Operator action:
See Section 4.2.4
“Attach the arterial blood line”
on
page A:79 in Operators handbook and press
Confirm
to
clear the attention.
Mute time:
2 minutes.