AK 98 Dialysis Machine - Attentions
B:81
506 BiCart is not attached to holder
506 BiCart is not attached to holder
Attach BiCart cartridge.
Appears:
When the BICART cartridge is not properly attached to the
holder.
Machine actions:
After 2 minutes there is a buzzer sound.
Operator action:
Attach or check BICART cartridge. See Section 4.2.3
handbook.
Mute time:
Permanent.