AK 98 Dialysis Machine - Measuring blood pressure
HCEN12745 Revision 10.2016
Program version 2.xx
CAUTION!
If the patient has other monitoring devices on the same arm as the cuff, these
devices may not work as they should while the blood pressure is measured.
CAUTION!
Do not apply a cuff to a limb that has restricted blood flow as this may affect the
blood pressure measuring.
CAUTION!
Do not apply a cuff to a limb with an access, for example a fistula or a graft as this
may have a negative effect of the fistula or graft.
8.3
Direct blood pressure measuring
Procedure
1)
Press the
BPM
readout field.
The measurement starts. The
heart in the read out field starts
flashing and BPM is flashing in the
status bar.
NOTE!
First inflation pressure is 180
mmHg.
Results
The measured value appears in the BPM read out field.
To stop measuring: Press the
BPM
read out field to stop an ongoing measurement.
8.4
Interval blood pressure measuring
Procedure
1)
Press the
BPM
button.
2)
Select the
Setup
tab.
3)
Press
Auto mode
.
4)
Press
Interval time
.
5)
Set how often the blood pressure monitor
should measure blood pressure, a value
between 5 and 60 minutes.
6)
Press
OK
.
7)
Press
Activate
to activate interval blood
pressure measurement.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...