Galvin Engineering CliniMix CP-BS WM-TMVPMB Product Installation Manualline Download Page 1

 

 

Version 2, 29 May 2020, 1 of 18 

Version 2,  26

th

 Feb 2020, 1 of 18 

CliniMix

®

 

CP-BS Hob Mounted Thermostatic 

Progressive Basin Mixer - Lever

 

 

PRODUCT CODES:

 

-    WM-TMVPMB 

 

TABLE OF CONTENTS 

1.0

 

INTRODUCTION 

1

 

2.0

 

SAFETY 

2

 

3.0

 

DESCRIPTION 

2

 

4.0

 

DIMENSIONS 

2

 

5.0

 

WATER SUPPLY CONDITIONS 

3

 

6.0 

 

INSTALLATION 

4

 

7.0

 

COMMISIONING 

6

 

8.0

 

OPERATION 

8

 

9.0

 

SERVICING 

9

 

10.0

 

MAINTENANCE 

11

 

11.0  

 

DISINFECTION 

13

 

12.0

 

WARRANTY 

15

 

13.0

 

TROUBLESHOOTING 

16

 

14.0

 

APPENDIX 

17

 

 

 

1.0 

INTRODUCTION 

The CliniMix

®

 CP-BS Hob

 

Mounted Thermostatic Progressive Basin Mixer - Lever 

is a high-performance 

thermostatic mixing valve suitable for a wide range of applications. The valve is designed to comply with 
‘AS4032. Thermostatic Mixing Valves - Materials, Design and Performance Requirements’. The mixing 
valve has the following features: 

 

Complies with the requirements of AS4032.4 – Thermostatic Mixing Valves. 

 

Provides high stability of mixed water temperature even under changing inlet conditions. 

 

Ensures rapid shut down of mixed outlet flow in the event of hot or cold water supply failure. 

 

Designed for quick and simple in-situ servicing. 

 

Suitable for installation into AS3500 compliant systems with hot water temperature as low as 55°Celsius. 

 

Fitted with tamper resistant temperature adjustment mechanism. 

 

Cool touch, non-conductive design in the hot side of the body, which minimises the risk of burns. 

 

Round smooth design to facilitate easy cleaning. 

 

Self-draining spout design. 

 

Hot water disinfection available without removing tap. 

IMPORTANT

: All GalvinCare

®

 mental health taps are tested in accordance with AS/NZS 3718 and leave 

our premises in good working order. Document to be read in conjunction with CliniMix

® 

Progressive 

Thermostatic Mixing Valve manual.

 

Summary of Contents for CliniMix CP-BS WM-TMVPMB

Page 1: ...g valve has the following features Complies with the requirements of AS4032 4 Thermostatic Mixing Valves Provides high stability of mixed water temperature even under changing inlet conditions Ensures...

Page 2: ...utions should be taken to ensure that these surfaces cannot cause such injuries 3 0 DESCRIPTION This manual covers the CliniMix CP BS Hob Mounted Thermostatic Progressive Basin Mixer Lever These are d...

Page 3: ...ecessary to remove the combined check valve and flow regulators and replace them with simple check valves Supply Water Temperature Range C Min 5 Max 85 Maximum Static Pressure kPa 1000 Working Pressur...

Page 4: ...ons It is important that both inlet dynamic supply pressures are 500kPa or less If either supply pressure exceeds 500kPa then a suitable pressure reducing valve must be fitted prior to the inlet contr...

Page 5: ...tap 4 Turn the mixer on and test operation Commission as per instruction 6 2 HANDLE POSITIONING 1 Fully open valve anti clockwise 2 Remove the handle assembly To prevent damage carefully prise off pla...

Page 6: ...re reading at the outlet with a digital thermometer The flow rate should be at least 4L min The temperature should be taken close to the mixer s outlet and if the outlet temperature requires adjustmen...

Page 7: ...flow should quickly reduce to be less than 0 4L min following the isolation The volume of mixed water discharged for a period of between 5 35 seconds should be measured and should not exceed the maxi...

Page 8: ...pplied with each TMV will ensure this requirement is met If the CliniMix Progressive Thermostatic Mixing Valve is to be installed without the supplied inlet control valves then it will be necessary to...

Page 9: ...ervicing checks should be carried out more frequently e g every 4 months Follow the recommended maintenance procedures detailed in section 10 During servicing note the following 1 Repeat the procedure...

Page 10: ...y with a 36mm AF deep socket Clean all components thoroughly and inspect cartridge for damage If the components are damage the cartridge must be replaced 5 The cleaned cartridge or a new cartridge if...

Page 11: ...commissioned as per instructions 10 0 MAINTENANCE When installed as an AS NZS 4032 4 application it is a requirement that the commissioning section 7 0 and maintenance procedures in section 10 are car...

Page 12: ...sis the anti splash nozzle should be inspected and cleaned if necessary This should be done immediately before sterilising is carried out In areas where lime scale build up is prevalent this will have...

Page 13: ...ote Care should be taken when carrying out the following procedure to avoid contact with hot water and hot surfaces We recommend the use of protective hand wear 1 Isolate water supply Turn both handle...

Page 14: ...es 30 Minutes 65 degrees 15 Minutes 70 degrees 10 Minutes Table 4 Guide to disinfection temperatures v s time To avoid scalding burning keep clear of the mixed outlet 6 Cool Down Unit This unit become...

Page 15: ...at system must be in accordance with its manufacturer s recommendation the Code and all relevant statutory and local State or Territory requirements 4 The valve must be returned to Galvin Engineering...

Page 16: ...r is damaged Non return devices are damaged Raise hot water temperature Clean the cartridge ensuring that all debris is not jammed Clean if necessary Mix temperature unstable Flow rate below 2L min Th...

Page 17: ...tails are to be filled in Incomplete reports will not be accepted PRINT ALL DETAILS or MARK WITH AN X IN BOXES TO INDICATE CHOICE Name of Facility Street Address Suburb State Post Code Phone Contact N...

Page 18: ...replaced and part numbers during this visit Service Kit No Other Parts Temperature range of warm water at outlet Neonatal and children 38 40 C Adult 40 5 43 5 C Set Temperature C Date at this service...