Figure 5.16: Alarms Display Configuration (Display Configuration)
The Alarms Display parameters, their options, and their functional
descriptions are given in Table 5.3.
Parameter Options
Functional Description
Line
1,2,3,4
Selects the line of the Transmitter LCD on
which the alarm for the selected parameter
is displayed.
“
---
“ indicates unselected (not displayed)
.
Display
Method
Steady
The parameter in Alarm will remain
steadily displayed on the given configured
line until the alarm is cleared.
Blinking
The parameter in Alarm will blink on the
given configured line until the alarm is
cleared.
Alternating
The parameter in Alarm will alternate
display with any other parameter selected
for display on the same line.
<Blank>
No indication of an alarm will be displayed
on the selected line if no alarmed
parameter is configured.
Lower
Limit
Numerical
Value
Lower Alarm Limit below which a
parameter goes into alarm. Expressed in
the units selected in the Units Display
submenu.
Upper
Limit
Numerical
Value
Upper Alarm Limit above which a
parameter goes into alarm. Expressed in
the units selected in the Units Display
submenu.
Table 5.3: Alarm Settings Parameters (Display Configuration)