3.4 Installation Requirements
Electrical Requirements
The power requirement is 10W.
Selecting the Installation Location of the
Transducer
3.4.2.1 Sample Temperature and Environmental
Considerations
The operating temperature ranges and
cooling
requirements
for the VL800 transducer are given in Section
3.4.2.2 Vibration Considerations
While the ViscoSite Viscometer can operate with some
process vibration, minimizing vibration at the installation
point will provide the best results. Figure 3.3 shows the
vibration tolerance for the VL800 viscosity transducer.
Figure 3.3: Vibration Tolerance Graph
The installation location should be checked with a vibration
analyzer that can determine both the frequency and
deflection of the processing line/tank vibration. These
measurements must be made under normal operating
conditions with all normally operating equipment running
to ensure representative data.
Sensor Oscillation Frequency:
400-700 Hz