ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 121 of 132
Data Type
Explanation
Allowed Values
mg
/cm
3
Stores entered manual
density value.
Entered as
mg
/cm
3
, not
g/cm
3
so that value is stored
as an integer number.
(Default=1000 mg/cm
3
=
1.000 g/cm
3
)
Milliseconds
Stores selected time
average window for time
averaged viscosity
parameter calculations
0 or positive integer
Position 3 Fixed-
Decimal
Stores selected Alarm Min
and Max Values for all 7
configurable alarms.
Stores selected Analog
Output Min and Max
values for Analog Outputs
Stores Calibration
frequency and calibration
temperature data.
Stores Temperature
Compensation variables.
Stores Relay Min and Max
values.
0 or any positive integer
where the last three digits of
the entered value are to the
right of the decimal point
(E.g. 6000 = 6.000, 50 =
050=0.050)
Relay Trigger
Stores selected relay
behaviour on alarm.
(Energized on alarm =>
Relay NO)
(De-energized on alarm =>
Relay NC)
0
De-energized (Open
on alarm)
1
Energized (Closed on
alarm)
Temperature
Units
Stores selected
temperature units
0
Degrees Celsius
1
Degrees Fahrenheit
Viscosity Units
Stores selected viscosity
parameter units
0
cP x g/cm
3
1
mPa
∙
s
2
cP
3
cSt
Table 11.1: Modbus Data Types, Descriptions, and Allowed Values