5.18.1 Schedule
The Schedule for each timer consists of the
Day
field where any specific day or “every day”
can be selected from the pull down menu and the
Hour
and
Minute
fields. The time fields
operate on 24 hour military time (e.g. 3:00AM=0300, 3:00PM=15:00).
Basically the analyzer cannot execute any timed gas runs scheduled from
12:00 AM to 12:59AM. in this mode Normally the time 00:15 would signifiy
15 minutes past midnight but for this application putting a zero in the hour
field would tell the analyzer to do whatever is scheduled in that column to
run every hour.
5.18.2 Analysis
The
Enable
checkbox must be checked before any of the gas runs listed below can be
executed on a timer. For any of the analysis types, the schedule time would be used to
initiate a single operation which may consist of one run or as many runs indicated in the
Consecutive
check box located for that run type on the
Streams
tab.
5.18.3 Sequences
The operations in this section are
Manual Sequences
that can be set to run for any specific
time period up to a week. The manual sequencers require a
Start time
and a
Stop time
:
therefore the date and time on two timers need to be entered. The
Start
and
Stop
checkboxes also need to be checked to enable this operation to execute.
Summary of Contents for ProTech903
Page 2: ......
Page 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Page 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Page 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Page 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Page 190: ...Revision 14 190 August 02 2019...