Section 3
Installation
3.1 Receiving the System
When the ACCUCHROME Natural Gas Chromatograph arrives, inspect the packaging for
external signs of damage. If there is any obvious physical damage, immediately contact the
shipping agent and Galvanic Applied Sciences to report the damage and request that the
carrier's agent be present when the unit is unpacked. It is recommended that you retain the
shipping container so that it may be used for future shipment of the unit, if necessary.
3.2 Environmental Requirements
3.2.1 Temperature/Humidity
The ACCUCHROME Natural Gas Chromatograph is designed to be operated at ambient
temperatures from -18
o
C to +60
o
C. The humidity should be between 0 and 95%, non-
condensing. For maximum accuracy and reliability it is recommended that the analyzer be
housed in an environmentally controlled shelter or enclosure.
For outdoor installations, the analyzer should be protected from direct sunlight and rain.
Enclosures are available from Galvanic Applied Sciences.
3.2.2 Space Requirements /Weight
The size and weight of the ACCUCHROME Natural Gas Chromatograph is presented in
Table
3-1. The installation site should provide adequate room for opening the cabinet doors for
maintenance and repair procedures. Complete dimensional information is provided in
Figures 3-1 and 3-2.
When the system is installed, leave 6” between the unit and other devices.
Table 3-1: Space/Weight of the ACCUCHROME Chromatograph
Size
Weight
Class 1, Div 2
33” H x 27”W x 11” D
85 lb
Class 1, Div 1
33” H x 27”W x 11” D
120 lb
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...