Revision 4
90
April 2020
Table 25: Modbus Communication Parameters
Parameter
Explanation
IP Port Number
Port number to be used if the communication protocol being
used is Modbus TCP/IP.
Modbus Address
Modbus address of the analyzer. This address needs to be
entered into the DCS to allow the DCS to communicate
correctly with the analyzer.
32 Bit Register Swap
If register swap is selected, the portion of the 32-bit register
that follows the decimal point is transmitted in the first 16-bit
word, and the portion that precedes the decimal point is
transmitted in the second 16-bit word. If register swap is not
selected, the portion that precedes the decimal point is
transmitted first.
NOTE: Only used if Modicon-32 format is chosen
Mode
Sets the mode for beginning and ending messages.
ASCII
mode uses ASCII characters to mark the beginning and
ending of messages, whereas
RTU
uses time gaps to mark
the beginning and ending of messages. Choice will depend
on the DCS being used.
Endian
Selects between
Big Endian
and
Little Endian
. In Big
Endian, the most significant bit is transmitted first, whereas
in Little Endian, the least significant bit is transmitted first.
Note: Only used if Modicon-32 format is chosen
Baud Rate
Selects the Baud Rate to be used for communication
between the analyzer and the DCS. Usually 9600 but may
differ depending on the DCS being used.
Note: Only used if RS485 serial communication is used
for Modbus
Data Bits
Number of data bits transmitted at a time. Usually 8, but 7
may be used under some circumstances
Note: Only used if RS485 serial communication is used
for Modbus
Parity
Type of parity checking used for data transmission. Usually
None
, but
Even
or
Odd
may be used under some
circumstances.
Note: Only used if RS485 serial communication is used
for Modbus
Stop Bits
Number of bits marking the end of a transmitted byte. Usually
1 but may be 2 under some circumstances.
Note: Only used if RS485 serial communication is used
for Modbus
The various items in the
Available Points
listing on the left side of the Modbus page can be
accessed by clicking on the + sign. The items in this listing can be used to populate the
Modbus Items
list on the right side of the Modbus page by dragging and dropping them into
the Modbus list. The general types of data in each folder in the
Available Points
list is given
in Table 26.
Summary of Contents for 943-TGX-CE
Page 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Page 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Page 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Page 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Page 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Page 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...