Revision 4
80
April 2020
Figure 49: Indicators Page
The specific faults shown on the Indicators page are listed in Table 20.
Table 20: Analyzer Faults
Fault
Explanation
Measure Out of Range
Spectrometer signal is out of range while analyzer is
measuring sample (signal either off scale or too low)
Reference Out of Range
Spectrometer signal is out of range while analyzer is in back
purge mode (signal either off scale or too low)
Dark Out of Range
Spectrometer dark level (i.e. signal when no light is present) is
either too low or too high
Low Cell Temperature
Measurement cell temperature is below low cell temperature
fault setpoint
High Cell Temperature
Measurement cell temperature is above high cell temperature
fault setpoint
Low Probe Temperature
Temperature of gas exiting the probe is below the low probe
temperature setpoint
High Probe Temperature
Temperature of gas exiting the probe is above the high probe
temperature setpoint
IO Board
Control computer is unable to communicate with the IO board
or spectrometer
Some of these faults may clear by themselves but others may require maintenance or even
hardware replacement to resolve. Refer to Section 7.2 for more detail on how to troubleshoot
active faults.
Table 21 describes each of the warnings listed int the warning table on the
Indicators
page.
Summary of Contents for 943-TGX-CE
Page 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Page 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Page 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Page 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Page 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Page 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...