Revision 5
48
August 2021
condition under which the Status relay can be in Fault while the Mode relay is in
Calibrate is if the analyzer power is off.
3.5.2.4
Control
The
Control
output indicates whether the analog output signals are valid. Table 6
describes the Control relay.
Table 6: Control Relay States
Condition
Display Indicator Colour
Explanation
Auto
Green
ALL the following conditions are true:
•
Status indicator is Normal (i.e. no
fault conditions present)
•
Mode indicator is in Run state
(i.e. analyzer is not performing a
zero calibration)
•
Online / Offline switch is in the
Online position (i.e. analyzer is
not under maintenance)
•
Analyzer power is on
Manual
Yellow
At least one of the conditions for the relay
to be in the Auto state is NOT true.
When the relay is in the
Auto
state, the analyzer is indicating that it believes that the data
being output on the analog outputs are valid current measurements of the tail gas process
steam. If the Control relay is in the
Manual
state, this indicates that the analyzer believes
that the data being output may not be valid. The purpose of this indicator is to signify
when the analyzer outputs are suitable for use as closed loop control inputs. When this
indicator is in the Manual state, no control action should be taken based on the analyzer
output signals. This indicator is generated on the basis of the other analyzer indicators,
as well as the condition of the
Online / Offline
mode toggle on the display (refer to
Section 4.3.1). The Control relay state is automatically returned to Auto once all the
conditions required for the Auto state have been met. While the Control relay is in the
Manual state, any analog output that is configured as Hold
will be held.
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...