Revision 5
39
August 2021
3.3.2 Disable System Bypass – Area Safe (Non-Explosive)
Once maintenance work is complete, and the installation environment remains safe (non-
explosive), the System Bypass should be disabled following the procedure below. This
procedure will prevent an automatic shut-down from occurring once the System Bypass is
disabled.
1. Close and latch the control cabinet door. Ensure that the SAFE PRESSURE indicator
above the EPCU display illuminates in blue and the ALARM / FAULT indicator stops
flashing red. If SAFE PRESSURE does not illuminate, disabling the System Bypass
will result in an auto-shutdown of the analyzer power supply.
2. Using the arrow keys, select BYPASS ON and press START/SET.
3. Using the arrow keys, select OFF and press START/SET. SETTING CORRECT? will
be displayed.
4. Select Y and press START/SET. The SYSTEM BYPASS indicated on the EPCU
display will turn off, and the SYSTEM BYPASS indicator will stop flashing yellow. The
EPCU is now once again in full-auto mode.
5. Press the Esc key twice to return to the EPCU’s default display.
3.3.2 Disable System Bypass – Area Hazardous (Explosive)
If the installation environment should suddenly become hazardous while maintenance work
is ongoing, the analyzer power must be immediately switched off with the disconnect device
mentioned in Section 2.4.3 OR the System Bypass immediately disabled.
1. Using the arrow keys, select BYPASS ON and press START/SET.
2. Using the arrow keys, select OFF and press START/SET. SETTING CORRECT? will
be displayed.
3. Select Y and press START/SET. The SYSTEM BYPASS indicated on the EPCU
display will turn off, and the SYSTEM BYPASS indicator will stop flashing yellow. The
EPCU is now once again in full-auto mode.
4. Press the Esc key twice to return to the EPCU’s default display.
Disabling the System Bypass in this fashion will cause an automatic shut-down, as safe
pressure will not be present in the control cabinet when the System Bypass is disabled.
3.4 Flow Control Settings
The flow control settings govern system operation. There are three instrument air flow rate
adjustment valves on the exterior of the oven cabinet, as shown in Figure 16.
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...