5-8
User Manual
Screen 5-4. Export Report Screen
CAUTION.
UseonlyaGalilMedicalsuppliedUSBflashdrivewiththeVisual-ICEMRISystem.
DonotusethesuppliedflashdriveforpurposesunrelatedtoVisual-ICEMRISystemdataand
reports.
4. Press the
Export
buttontobeginexportingthefile.Waitforconfirmationbeforeremovingthe
USBflashdrivefromthesystem.
Screen 5-5. Report Exported message
5.4 Changing Gas Cylinders during a Procedure
If it becomes necessary to replace a gas cylinder during a procedure, stop all freezing and thawing
operations.
5.4.1 Standard Gas Cylinder Setup
1. Plan the appropriate time to make a cylinder change by estimating the gas required to complete
the procedure. The
Gas Indicator
on the
Navigation Tool Bar
indicates how much time is
remainingineachgascylinderbasedontheselectedgasflowintensity,typeandnumberof
needles being used. Also, consider the number of planned freeze-thaw cycles for the procedure.
2. Safely position a full gas cylinder, with the required gas type and purity, near the empty cylinder.
3. Close and tighten the cylinder valves of both gas cylinders.
4. Slowly open the Manual Vent Valve to discharge gas from the system and the high-pressure
gas supply line. Wait until all the pressure is released and both gauges on the gas supply lines
indicate zero pressure.
5. Use the appropriate wrench to remove the gauge assembly from the empty cylinder.
6. Connect the gauge assembly to the full cylinder.
Summary of Contents for Visual-ICE MRI
Page 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Page 20: ...1 10 User Manual This page intentionally left blank...
Page 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Page 56: ...4 14 User Manual This page intentionally left blank...
Page 66: ...5 10 User Manual This page intentionally left blank...
Page 80: ...9 2 User Manual This page intentionally left blank...
Page 118: ...12 2 User Manual This page intentionally left blank...
Page 120: ...13 2 User Manual This page intentionally left blank...
Page 121: ......