Table 11-2. Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The Visual-ICE MRI Cryoablation System is intended for use in a professional healthcare facility
electromagnetic environment with the compliance levels specified below. The customer or the user
of the Visual-ICE MRI System should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The Visual-ICE MRI System uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class A
Harmonic emissions, IEC
61000-3-2
Class A
Voltage fluctuations/ flicker
emissions, IEC 61000-3-3
Complies
NOTE:
The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is
normally required) this equipment might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or reorienting the
equipment.
Summary of Contents for Visual-ICE MRI
Page 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Page 20: ...1 10 User Manual This page intentionally left blank...
Page 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Page 56: ...4 14 User Manual This page intentionally left blank...
Page 66: ...5 10 User Manual This page intentionally left blank...
Page 80: ...9 2 User Manual This page intentionally left blank...
Page 118: ...12 2 User Manual This page intentionally left blank...
Page 120: ...13 2 User Manual This page intentionally left blank...
Page 121: ......