Gage CompuScope 1220 Hardware Manual Download Page 9

CompuScope 1220 

 

CompuScope 1220 specifications 

PLEASE CHECK THE GAGE WEBSITE FOR THE MOST UP-TO-DATE SPECIFICATIONS. 

SYSTEM REQUIREMENT 

PCI-based computer with at least one free full-length PCI slot, 128 MB RAM, 50 MB hard disk and SVGA 
video. 

SIZE 

Plugs into one full-length PCI slot, 13” x 4.1”. 

1 M Acquisition Memory 

1 slot 

4 M Acquisition Memory 

2 slots 

8 M Acquisition Memory 

2 slots 

64 M Acquisition Memory 

3 slots 

256 M Acquisition Memory 

3 slots 

512 M Acquisition Memory 

3 slots 

1 G Acquisition Memory 

3 slots 

POWER (IN WATTS) 

+ 5 Volts 

Acquisition 

Memory Worst 

Case 

Typical 

1M 

25.0 W 

17.5 W 

4M 

28.0 W 

20.5 W 

8M 

28.0 W 

20.5 W 

64M 

32.5 W 

23.5 W 

256M 

32.5 W 

23.5 W 

512M 

32.5 W 

23.5 W 

1G 

32.5 W 

23.5 W 

Note:  Power connector on the deep memory board of 64M, 256M, 512M and 1G models must also be connected 

using a Y-cable. 

CHANNELS A & B 

Inputs per card:   

2 single-ended inputs 

Impedance:  

M

, 35 pF or 50 

, software selectable 

Coupling:   

 

AC or DC 

Connector: 

 

1 BNC per channel 

Resolution:  

12 

bits 

A/D Type: 

 

Monolithic, 16 bit oversampling with decimation filter, operating in 12 bit mode 

Input Voltage Ranges: 

± 50 mV, ± 100 mV, ± 200 mV, ± 500 mV,  
± 1 V, ± 2 V, ± 5 V, ± 10 V 

Analog Bandwidth: 

Input Range 

-3dB Point 

± 10 V 

5 MHz 

± 5 V 

5 MHz 

± 2 V 

10 MHz 

± 1 V 

10 MHz 

± 500 mV 

10 MHz 

± 200 mV 

10 MHz 

± 100 mV 

10 MHz 

± 50 mV 

10 MHz 

Summary of Contents for CompuScope 1220

Page 1: ...GAGE APPLIED TECHNOLOGIES CS1220 Hardware Manual Reorder MKT HWM PCI01 CS1220 0408...

Page 2: ...The installation program used to install the GageScope Software InstallShield is licensed software provided by InstallShield Software Corp 900 National Parkway Ste 125 Schaumburg IL InstallShield is C...

Page 3: ...20 4 CompuScope 1220 compliance statement 7 CompuScope 1220 product introduction 8 CompuScope 1220 specifications 9 CompuScope 1220 ordering information 13 CompuScope 1220 simplified block diagram 14...

Page 4: ...ompuScope 1220 card If you purchased a CompuScope 1220 4M or CompuScope 1220 8M card you should see a piggyback memory board attached to the back of the card as shown below If you purchased a CompuSco...

Page 5: ...e Note that some packages will only be available if you have purchased the software and have a key provided by Gage CompuScope Certificate of NIST Traceable Calibration Each CompuScope card is shipped...

Page 6: ...er Slave Timing Module is used for 4 Card Master Slave system of CS1220 1M 2 Card Master Slave system of CS1220 4M or CS1220 8M 8 Slot Master Slave Timing Module 8 Slot Master Slave Timing Module is u...

Page 7: ...l Fast Transient Burst Immunity Performance criterion B IEC61000 4 5 Power Line Surge Immunity Performance criterion B IEC61000 4 6 Conducted RF Immunity Performance criterion A IEC61000 4 11 Voltage...

Page 8: ...es from CompuScope memory to PC memory run as high as 50 MB s for the CS1220 as compared to a few hundred KB s for GPIB CompuScope cards are easier to program as Software Development Kits are availabl...

Page 9: ...ts Acquisition Memory Worst Case Typical 1M 25 0 W 17 5 W 4M 28 0 W 20 5 W 8M 28 0 W 20 5 W 64M 32 5 W 23 5 W 256M 32 5 W 23 5 W 512M 32 5 W 23 5 W 1G 32 5 W 23 5 W Note Power connector on the deep me...

Page 10: ...of 95 of full scale on the 1 V range Typical values listed below SNR 55 dB SFDR 57 dB ENOB 8 9 bits ACQUISITION MEMORY Data Storage In on board memory Memory sizes 1 Msamples 4 Msamples 8 Msamples 64...

Page 11: ...a 64 point resolution Maximum Number of Triggers 4 194 304 MULTI CARD SYSTEMS Operating modes Master Slave or Multiple Independent Maximum Number of Cards Master Slave 2 to 8 cards for 1M memory 2 3 o...

Page 12: ...nced analysis tools except for Extended Math Professional Edition Provides full functionality of all advanced analysis tools SOFTWARE DEVELOPMENT KITS SDK CompuScope SDK for C C For Windows 98 ME NT 2...

Page 13: ...1 003 GageScope Software Product Order No GageScope Lite Edition Included GageScope Standard Edition purchased with CompuScope hardware 300 100 351 GageScope Standard Edition purchased independently 3...

Page 14: ...14 CompuScope 1220 CompuScope 1220 simplified block diagram Figure 1 CS1220 simplified block diagram...

Page 15: ...ad of the CompuScope card s own oscillator External Trigger BNC Connector used to input an analog or digital signal which may be used as an External Trigger External Trigger is defined exactly as in a...

Page 16: ...note that much higher PRFs will be achieved using CompuScope Multiple Record mode The following test results were obtained using a computer configured as follows Pentium III 1 GHz processor 512 MB RAM...

Page 17: ...CompuScope 1220 17 Notes...

Reviews: