CompuScope 1220 throughput & maximum PRF 

A number of applications require the CompuScope 1220 to acquire data based on a rapidly occurring trigger 
signal.  These high Pulse Repeat Frequency (PRF) applications include imaging, radar, ultrasound and lightning 
test. 

Gage has performed extensive repetitive capture benchmarks in single record mode.  In this mode, the signal is 
captured into on-board CompuScope memory and the captured data are transferred through the PCI bus using 
PCI bus mastering to PC RAM.   

Please note that much higher PRFs will be achieved using CompuScope Multiple Record mode. 

The following test results were obtained using a computer configured as follows: 

• 

Pentium III, 1 GHz processor 

• 

512 MB RAM 

• 

20 GB disk drive 

• 

Windows 2000 

• 

NT File System 

• 

33 MHz, 32 bit PCI bus 

• 

All slots support bus mastering 

A C application program optimized for fast repetitive capture in single record mode was used for throughput 
measurements.  The CS1220 was operated using this application for many different capture depths and the 
results are plotted as points in the graph below.  The PCI transfer rate was calculated from the linear portion of 
the curve at high depths. 

PCI data transfer rate: 

50 MB/s 

 

Figure 3:  Maximum PRF vs. acquisition length

Summary of Contents for CompuScope 1220

Page 1: ...GAGE APPLIED TECHNOLOGIES CS1220 Hardware Manual Reorder MKT HWM PCI01 CS1220 0408...

Page 2: ...The installation program used to install the GageScope Software InstallShield is licensed software provided by InstallShield Software Corp 900 National Parkway Ste 125 Schaumburg IL InstallShield is C...

Page 3: ...20 4 CompuScope 1220 compliance statement 7 CompuScope 1220 product introduction 8 CompuScope 1220 specifications 9 CompuScope 1220 ordering information 13 CompuScope 1220 simplified block diagram 14...

Page 4: ...ompuScope 1220 card If you purchased a CompuScope 1220 4M or CompuScope 1220 8M card you should see a piggyback memory board attached to the back of the card as shown below If you purchased a CompuSco...

Page 5: ...e Note that some packages will only be available if you have purchased the software and have a key provided by Gage CompuScope Certificate of NIST Traceable Calibration Each CompuScope card is shipped...

Page 6: ...er Slave Timing Module is used for 4 Card Master Slave system of CS1220 1M 2 Card Master Slave system of CS1220 4M or CS1220 8M 8 Slot Master Slave Timing Module 8 Slot Master Slave Timing Module is u...

Page 7: ...l Fast Transient Burst Immunity Performance criterion B IEC61000 4 5 Power Line Surge Immunity Performance criterion B IEC61000 4 6 Conducted RF Immunity Performance criterion A IEC61000 4 11 Voltage...

Page 8: ...es from CompuScope memory to PC memory run as high as 50 MB s for the CS1220 as compared to a few hundred KB s for GPIB CompuScope cards are easier to program as Software Development Kits are availabl...

Page 9: ...ts Acquisition Memory Worst Case Typical 1M 25 0 W 17 5 W 4M 28 0 W 20 5 W 8M 28 0 W 20 5 W 64M 32 5 W 23 5 W 256M 32 5 W 23 5 W 512M 32 5 W 23 5 W 1G 32 5 W 23 5 W Note Power connector on the deep me...

Page 10: ...of 95 of full scale on the 1 V range Typical values listed below SNR 55 dB SFDR 57 dB ENOB 8 9 bits ACQUISITION MEMORY Data Storage In on board memory Memory sizes 1 Msamples 4 Msamples 8 Msamples 64...

Page 11: ...a 64 point resolution Maximum Number of Triggers 4 194 304 MULTI CARD SYSTEMS Operating modes Master Slave or Multiple Independent Maximum Number of Cards Master Slave 2 to 8 cards for 1M memory 2 3 o...

Page 12: ...nced analysis tools except for Extended Math Professional Edition Provides full functionality of all advanced analysis tools SOFTWARE DEVELOPMENT KITS SDK CompuScope SDK for C C For Windows 98 ME NT 2...

Page 13: ...1 003 GageScope Software Product Order No GageScope Lite Edition Included GageScope Standard Edition purchased with CompuScope hardware 300 100 351 GageScope Standard Edition purchased independently 3...

Page 14: ...14 CompuScope 1220 CompuScope 1220 simplified block diagram Figure 1 CS1220 simplified block diagram...

Page 15: ...ad of the CompuScope card s own oscillator External Trigger BNC Connector used to input an analog or digital signal which may be used as an External Trigger External Trigger is defined exactly as in a...

Page 16: ...note that much higher PRFs will be achieved using CompuScope Multiple Record mode The following test results were obtained using a computer configured as follows Pentium III 1 GHz processor 512 MB RAM...

Page 17: ...CompuScope 1220 17 Notes...

Reviews: