background image

 

G-Tek Corporation Pvt. LTD  

 

Introduction 

 

1.

 

I

NTRODUCTION

 

This manual is written to help the user to get familiar with the installation and operation of LM Pro RF. 
It is designed for reliable and convenient recording of process parameters in production, Research and 
Quality control applications. This device can sense the Temperature and Humidity, displaying them on 
LCD  display  ae  well  as  sending  these  data  via  RF  signal  to  Moderator  device  for  monitoring  on 
gtekNetTM.   

1.1

 

 

M

ANUAL 

L

AYOUT

 

This manual is divided into sections for quick and easy reference.  

Table 1 Manual Layout 

Section 1 Introduction 

This gives outline of the manual, brief description about the LM Pro 
RF, features available and how to unpack it. 

Section 2 Installation 

This section gives the details of the Electrical installation. 

Section 3 Operation 
 

This section describes the operation of LM Pro RF and the output 
status on the LCD display. 

Section 4 
Troubleshooting Guide 

This section describes about the most frequently asked questions 
and their solutions about LM Pro RF. 

Section 5 Accessories 

This section shows the standard accessories for the LM Pro RF. 
 

Section 6 Specification 

It describes the detailed specification of the LM Pro RF. 
 

Section 7 Ordering Code 

It  describes  the  details  of  order  code  and  comparing  it  with  the 
code on the LM Pro RF, user can find out the installed options.

 

 

1.2

 

 

LM

 

P

RO 

RF

 

D

ESCRIPTION 

 

 

The LM Pro RF series of wireless data logger is a new RF based data logger measuring temperature 
and % RH. These data loggers are very useful in monitoring data across large facilities where wiring is 
difficult. Since no wiring is required, deployment is very quick and easy. Data are sent by each device 
directly to the centralized moderator (repeater) device and from there to gtekNetTM software over 
ethernet port. These data can be accessed from any network connected device using standard web 
browser.  

The data logger consists of RF radio transmitter and receiver operating at 869.88 MHz (SRD License 
Free:  EU  Version)  frequency.  These  data  loggers  form  a  star  network  with  its  moderator.  The 
moderator is connected to the client-server based gtekNetTM software through ethernet. A new data 
logger when it is in the range of the central node 

 moderator, gets easily connected to the gtekNetTM. 

Each device transmits its data to the moderator and is captured in gtekNetTM. Any parameter of the 
data logger when set / changed from gtekNetTM, first goes to the moderator and then to the device. 
These data loggers have an internal memory of 32000 data sets. In the event of communication failure 
between moderator and device or moderator and gtekNetTM, after reconnection, all the missed data 
will be synchronized by gtekNetTM automatically. This ensures that no data is missed.  

Summary of Contents for LM Pro RF

Page 1: ...LM RF Temp RH Datalogger Model No 6xx G Tek Corporation Pvt Ltd 3 mahavir estate karelibaug vadodara 390 018 tel 91 265 2461912 email info gtek india com url www gtek india com OPERATING MANUAL LM H...

Page 2: ...s 5 1 4 Unpacking and inspection of LM Pro RF 5 1 4 1 Status LEDs 6 1 4 2 Display LCD 6 Installation 7 Operation 8 3 1 LM Pro RF Internal Schematics Diagram 8 3 2 LM Pro RF OPERATION 8 Troubleshooting...

Page 3: ...List of Figures Figure 1 Front View of LM Pro RF 5 Figure 2 LCD Display 6 Figure 3 Schematic diagram of Set up of LM Pro RF 7 Figure 4 Internal Schematic diagram of LM Pro RF 8 Figure 5 5 VDC 1 2 Amp...

Page 4: ...G Tek Corporation Pvt LTD 3 List of Tables LIST OF TABLES Table 1 Manual Layout 4 Table 2 Specification of LM Pro RF 11 Table 3 Sensor types and range 12 Table 4 LM Pro RF Order Code 13...

Page 5: ...nd comparing it with the code on the LM Pro RF user can find out the installed options 1 2 LM PRO RF DESCRIPTION The LM Pro RF series of wireless data logger is a new RF based data logger measuring te...

Page 6: ...INSPECTION OF LM PRO RF LM Pro RF is dispatched in a recyclable environment friendly package specially designed to give adequate protection during transit If the outer box shows sign of damage it shou...

Page 7: ...ll be flickered with Heartbeat LED if Batch is ON at every 5 minutes Otherwise Storage LED will remain in OFF state if Batch is OFF 1 4 2 Display LCD The LCD display is 3 digit display shows the Tempe...

Page 8: ...sing ethernet cable for communication Please refer the schematic diagram given in figure 3 for setting up the connection The PC must have GtekNet application to configure the Moderator for observing t...

Page 9: ...pdated The STR led will flicker with HBT Led at 5 minutes time interval if Batch is running Otherwise STR led will remain in OFF state When the Batch is running recording indication is shown on LCD di...

Page 10: ...LED will be flickered and Temp and RH will be updated If HRT LED does not flicker then contact to Gtek Corporation How is Alarm condition shown on LCD display When Alarm occurs it will be shown on LCD...

Page 11: ...ard Accessories STANDARD ACCESSORIES 5 1 5 V DC 1 2 AMP ADAPTOR Figure 5 5 VDC 1 2 Amp Adaptor Input Voltage 100 240 V AC 0 35 Amp Operating Frequency 50 to 60Hz Output voltage 5V DC 1 2 Amp 5 2 ETHER...

Page 12: ...ords totals Rollover Environmental Temperature Operation 5 C to 45 C Limiting 0 C to 50 C Storage 20 C to 60 C Humidity Operation 10 to 80 RH Non Condensing Storage 5 to 90 RH Non Condensing Altitude...

Page 13: ...t LTD 12 Specification Table 3 Sensor types and range Parameter Sensor Range Accuracy Temperature Thermistor 20 to 60 C 0 5 C Temperature RH Solid state Temp 20 to 60 C 0 5 C RH 0 100 2 RH between 20...

Page 14: ...er Code Series Input Type Power Supply Mounting Type Software Communication Interface 6 LM Pro RF 1 Temp 0 Battery Operated 0 Handheld 4 GtekNet Non Secure 6 Radio Frequency RF 2 Temp RH 1 Battery Mai...