![Fungilab EVO H Instruction Manual Download Page 88](http://html.mh-extra.com/html/fungilab/evo-h/evo-h_instruction-manual_2346306088.webp)
EVO EXPERT Manual
88/105
Appendix A:
Wireless Datalogger configuration
This Appendix aims to explain and guide the user through the configuration of the FTP system
incorporated in this viscometer model.
In order to use this feature, it is required to have an FTP server ready to be accessed. Additionally,
it is necessary to have a USB connection between the viscometer (in Fungilab Data Boss mode) and
a computer in order to configure the different parameters of the wireless connection.
A.1.-Wireless Parameters
The wireless capabilities require the configuration of several parameters. These parameters are
required to connect the viscometer to the wireless network, contact the FTP server and upload the
datalog saved in its memory.
The required parameters are presented below. Some are optional depending on the conditions
defined by the wireless network. This is indicated with an ‘[o]’ after the parameter and a little
explanation in italics after the description. Moreover, a pyramid chart that represents these
parameters in a hierarchical structure is shown in Fig.17.
FTP PARAMETERS
Server IP address: This parameter sets the IP value that identifies the FTP server on the
network.
Server port: This parameter sets the port through which the viscometer accesses the FTP
server.
FTP folder: This parameter sets the folder from the FTP root directory in which the file is
going to be saved. If left blank, the file is saved in the ‘PUBLIC’ folder of the root directory
(/PUBLIC). An error screen appears if this folder cannot be found or it does not exist. Put a
dot character (.) to save the file in the main root.
FTP file: This parameter sets the name of the file to be saved into the FTP server. This
name must also contain the file extension (.csv/.txt).
FTP username: This parameter sets the user of the FTP server that the viscometer will use
to access it.
FTP password: This parameter sets the user password for accessing the FTP server.
IP CONFIGURATION
Static/Dynamic IP address: Disables/Enables the DHCP for automatic IP configuration inside
the wireless network. I case of Static IP address, the following fields must be implemented:
o
IP address[o]: This parameter sets the IP address that identifies the Viscometer
inside the network. For the proper function of the communication, it is necessary to
have a unique IP address for the Viscometer.
o
Netmask[o]: This parameter sets the netmask that defines the IP groups inside the
network. The default value is usually 255.255.255.0.
o
Gateway[o]: This sets the IP address of the gateway that manages the wireless
network.
These three fields are not mandatory if the DHCP is enabled (dynamic IP address).
Otherwise, these fields are required to ensure the proper working communication.
WIRELESS PARAMETERS