5-8
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
reliability of the waveform. A red display indicates low
reliability and a green display indicates high reliability.
If the display remains red, press the [STOP] key to stop
the measurement and check whether the cuffs or PCG
microphone are attached correctly and whether the
patient is calm.
(Refer to "Real Time CAVI Check" (P5-12) .)
10
If "ON" is set for [PW/BP Exam.] - [Pulse
Exam. Set.] - [CAVI Check], the CAVI
Check display will appear after the CAVI
measurement is completed.
(Refer to "CAVI Check Result Display" (P5-12) )
11
Measure the right brachial and right
ankle blood pressure.
The pause duration can be set between 5 and 120
seconds in "Meas Pause Right to Left" in the BP Exam.
Settings.
12
Change the cuff sites. Remove the cuffs
from the right brachium/ankle and attach
them to the left brachium/ankle.
A measurement site guide is displayed. Attach the
cuffs in the indicated position. (Refer to "Attachment
Site Guide" (P5-11) .)
13
After the cuffs have been attached,
press the [START] key.
• Inflate the cuffs on the left brachium/ankle to acquire a
pulse wave.
• The CAVI Check display appears. (Refer to "CAVI
Check Result Display" (P5-12) )
14
Measure the left brachial and left ankle
blood pressure.
15
Upon completion, the examination
result will be displayed. (Refer to
"Examination Result Display" (P5-14) .)
If the color printer is set to automatic printing, the results
are automatically printed after the measurements are
completed.
■
If "ON" is set for [File] - [General] -
[Auto Save] in the Settings:
The data is automatically saved to the media set for
[Save Media].
16
Press the [Back] button.
The display will return to the home display.
The results can be checked before they are automatically
saved or sent. This mode is enabled if "ON" is set for [PW/
BP Exam.] - [Pulse Wave: Gen] - [Confirm results] in the
Settings. Touch [Back] in the result display to return to the
home display without saving or sending. (In this case, the
measurement results will not be saved.)
If CAVI cannot be measured due to
causes such as noise interference, the
message "Cannot measure CAVI..."
appears in the CAVI Check display and
the measurement will be suspended. To
stop the measurement and perform the
measurement again, press the [STOP]
key. To continue measuring, touch
[Cancel] or press the [START] key. The
blood pressure measurement will
proceed.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...