5-21
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
3
Enter the distance between the
measurement sites.
1)
Touch the button corresponding to the site
to enter a value.
2)
Enter the value.
3)
Touch [
].
• The L1 distance is automatically calculated
and entered when the AF distance is entered.
• The total L distance is calculated when all
values are entered.
4)
Repeat steps 1) - 3) to enter all necessary
values.
5)
When all values have been entered, touch
[
] in the "Section" window.
●
Entering Segment Values
1
Measure the distance between the
measurement sites.
Place the tape measure with the "0" point (starting
point) at the center of the right ankle cuff (point D in
the diagram) and measure the distances from there to
the center of the right knee joint (point C in the
diagram), the pulse position of the right femoral artery
(point B in the diagram) and the left side of the left
second intercostal sternum (point A in the diagram).
Knee Joint
Scale reading (cm) at point C.
Femoral Artery
Scale reading (cm) at point B.
Aortic Valve
Scale reading (cm) at point A.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...