6-3
Chapter 6 Admit/Discharge
Admit
7
Patient Classification [Adult]/ [Child]/ [Neonate]
The patient classification selection will affect the accuracy of NIBP, HR, RR measurement. It will also affect the
delay time of numeric data alarm.
The selected patient classification and icon will be displayed on the home display.
WARNING
The patient classification selection influences the precision of the QRS detection and
NIBP measurement. Make sure the correct selection is made.
NOTE
Under [Initial Settings > User I/F > Admit > Alarm], alarm settings at admittance can be
set to link to patient classification.
8
Alarm at Admittance
By selecting the preprogrammed alarm mode, the current alarm settings can be changed.
9
Sex [Male] / [Female]
10
Nurse Team
To easily recognize the patients for each nurse team, the displayed colors on the home display and individual
bed display can be changed according to the nurse team.
The nurse teams need to be registered in advance on the "Initial Settings".
(
11
Pacemaker
Select from [Used] / [Not Used].
WARNING
The pacemaker use selection influences the precision of the QRS detection and
arrhythmia analysis. Make sure the correct selection is made.
When [Used] is selected, the monitor will identify the pacemaker pulse and insert an artificial pulse onto the
ECG waveform for easy identification. When pacing waveform does not appear (pacing failure), erroneously
detecting the pacemaker pulse as QRS will be prevented.
The arrhythmia analysis will detect pacing beat as P (Pacemaker Beat) or F (Fusion Beat) to prevent
erroneous judgment of VPC.
12
Comment
A comment can be entered.
The entered comment can be displayed on the patient information area by selecting [Comment] for "Disp.
Item for Patient Info. Area" under [Common Setup > Display Config. > Detail Setup > Other].
Pressing the [Fixed Form] key on the "Comment" window will display the comments preprogrammed on
"Register Fixed Comment" under [Initial Settings > User I/F > Display/Print]. (
Maintenance Manual
"Display/Print" P7-12)
Pressing the key next to "Sort" will sort the fixed comments by colors preprogrammed on "Register Fixed
Comment".
For adult:
For child:
For neonate:
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......