12-14
Chapter 12 Printing
Measurement Status
It will stack a new data instead.
Measurement Status
Measurement status such as vital signal condition and equipment status will be printed after the printing type.
The following measurement status will be printed.
Measurement Status
Description
LEAD OFF
The electrode is detached. Check electrodes.
CVA
CVA is detected.
P SEARCH
SpO
2
pulse wave is small.
SpO
2
probe sensor attachment is not appropriate, etc.
ECG LOW
The amplitude of ECG waveform is too low.
ECG CAN'T
Cannot analyze
ECG1 LOW
The amplitude of ECG1 waveform is low.
ECG2 LOW
The amplitude of ECG2 waveform is low.
ECG1 CAN'T
Cannot perform arrhythmia analysis of ECG1.
ECG2 CAN'T
Cannot perform arrhythmia analysis of ECG2.
CO
2
CHECK
CO
2
sensor error.
TLM OFF
Too far, noise interference.
TLM LOWBAT
Telemetry battery or bedside monitor battery is depleted.
LAN OFF
Communication error with DS-LANIII.
DS-LANIII connection is cut off.
Cannot receive data via DS-LANIII.
TCON OFF
TCON signal cannot be received.
Data cannot be received by TCON.
TCON ERROR
Communication error with TCON.
CHECK SPIRO
SPIRO unit is malfunctioning.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......