8-12
Chapter 8 Parameter Setup
RESP
Common Setup / Impedance Setup
RR can be measured from the following 3 sources.
Impedance RR
RR measured on the multigas unit
RR measured on the ventilator
There are setup items common to all sources and setup items specific to each source.
Common Setup
1
RR Synchronized Mark
[ON]: The RR synchronized mark will be displayed.
[OFF]: The RR synchronized mark will not be displayed.
2
RR/APNEA Alarm Source
RR/APNEA Alarm Source can be set only on the bedside monitor. The setting will be displayed on the central
monitor.
[Impedance]: RR alarm will be generated based on the impedance respiration curve. Also, the RR
synchronized mark based on impedance respiration will be displayed.
[CO
2
/GAS]: RR alarm will be generated based on the RR measured by the CO
2
module or multigas unit.
Also, the RR synchronized mark based on CO
2
waveform will be displayed for only the wired network beds.
[Vent.]: RR alarm will be generated based on the RR measured by the ventilator. Also, the RR synchronized
mark based on ventilator measurement will be displayed for only the wired network beds.
Impedance Setup
1
CVA Detect
[ON]: When CVA is detected, alarm will generate and message will be displayed.
[OFF]: CVA detection will not be performed.
When the amplitude of the respiration waveform decreases due to causes such as respiratory pause, the ECG
waveform may be superimposed on to the respiration waveform, making the RR equal to the HR. This
condition is called CVA (Cardio-Vascular Artifact), and is detected using the CVA detection function
This function will be effective only when [Impedance] is set as the "RR/APNEA Alarm Source".
If the ECG waveform is superimposed on to the respiration waveform with HR (RR) of 30Bpm or above for
more than 20 seconds (10 seconds for neonates) and if the "CVA Detect" is set to [ON], the <CVA detected>
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......