8-7
Chapter 8 Parameter Setup
ECG
The following items can be set on the "ECG Detail Setup" screen.
1
Filter
For the bedside monitor, the waveform frequency characteristic can be selected from Monitor Mode, ESIS
Mode, or Diagnosis Mode according to the monitoring purpose. The settings will be transmitted to the central
monitor.
CAUTION
The ESIS mode can largely reduce the artifact such as electrosurgery noise and EMG,
but it may also reduce the QRS amplitude. Using the ESIS mode may erroneously detect
the pacemaker spike.
The ESIS mode should be selected only when a high frequency noise largely affects the
HR measurement.
This setup is not available for LW bed.
2
Synchronized Mark/Tone
[OFF]: Synchronized mark will not be displayed.
[ECG]: HR synchronized mark will be displayed. The synchronized
tone will be set to ON.
[SpO
2
]/ [SpO
2
-2]: SpO
2
synchronized mark will be displayed. The
synchronized tone will be set to ON.
[BP]: BP synchronized mark will be displayed. The synchronized tone will be set to ON.
3
Pacemaker
[Used]: Pacemaker pulse will be detected and pace pulse mask function will be performed for set
duration.
[Not Used]: Pacemaker pulse will not be detected.
4
Pacemaker Pulse
Pacemaker Pulse Detection Algorithm is as follows.
1 ECG Signal Input
ECG signal will be input.
2 Pacemaker Pulse Detection and Suspension of QRS Detection
Detects the high frequency and large amplitude signal as pacemaker pulse.
When pacemaker pulse is detected, QRS detection will be suspended for fixed
amount of time to avoid erroneous detection of pacemaker pulse as QRS.
3 Canceling of Arrhythmia Detection
Arrhythmia detection of the waveform following the pacemaker pulse will be
cancelled.
CAUTION
Precautions about Pacemaker Pulse Detection
There are some cases when the pacemaker pulse can not be detected depending on
the pacemaker type, pulse voltage, pulse width, electrode lead type (unipolar, bipolar),
First Page
Filter, Synchronized Mark/Tone, Pacemaker, Pacemaker Pulse
Second Page
Pace Pulse Mask Time, QRS Detection, ECG Drift Filter, AC Filter
Third Page
Auto Lead, ST/VPC/Arrhy. Alarm Display
2
1
3
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......