7-14
Chapter 7 Alarm Function
Alarm Silence and Alarm Sound Suspend
CAUTION
Pressing the [Alarm Silence] key will silence all the alarms generated on the displayed
beds. Pay attention not to miss any important alarms.
REFERENCE
When [Disable] is set for "All Beds Alarm Silence Key" (Initial Settings > Alarm), [Alarm
Silence] key for the fixed key, user key, remote control unit will be disabled.
(
Maintenance Manual "Alarm" P7-2)
2
To silence the alarm for specific bed, press the event key displayed
at alarm generation (shown on right), or [Individual Alarm Silence]
key on the individual bed display.
3
When an alarm is silenced on the bedside monitor, the alarm for that
bed will be also silenced on this equipment.
Precautions about Silencing the Alarm
If the alarm cause still remains at completion of alarm silence time (1min. / 2 min.), the alarm sound will
generate again.
If the [Alarm Silence] key is pressed while NIBP alarm is generating, NIBP alarm will be canceled and alarm
sound will not generate again even after the preprogrammed alarm silence duration.
If the [Alarm Silence] key is pressed for the alarm of another parameter which occurred during the alarm
silence condition, the alarm silence duration for the first alarm will not be extended.
If another alarm with the same or higher level occurs during the alarm silence condition, the alarm sound for
the new alarm will occur.
Depending on the "Alarm System" setting under "Initial Settings", the alarm operation will differ as follows.
To Cancel the Alarm Silence Condition
The alarm silence condition can be cancelled.
Pressing the [Resume All Al. Sound] key will cancel the alarm silence condition for the selected bed and resumes
the alarm sound if alarm factor exists.
If an alarm condition is resolved for a moment but is generated again during the alarm silence time:
When [Fukuda Tone] is set
Alarm sound will not generate. Recall and alarm printing will not be
performed.
When [Melodic Tone] or [Standard Tone] is set Alarm sound will generate. Recall and alarm printing will be performed.
If another alarm with lower priority occurs during the alarm silence time:
When [Fukuda Tone] is set
Alarm sound will not generate. Recall and alarm printing will be
performed.
When [Melodic Tone] or [Standard Tone] is set Alarm sound will generate. Recall and alarm printing will be performed.
If numeric data alarm or arrhythmia alarm with lower priority occurs during the equipment status alarm silenced duration.
When [Fukuda Tone] or [Melodic Tone]
or [Standard Tone] is set
Alarm sound will generate.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......