Chapter 7 Alarm Function
Alarm Occurrence
For each parameter, upper/lower alarm limit, current value, alarm OFF icon, nurse
call mark will be displayed.
Pressing the parameter area will display the floating window for the corresponding
parameter. (shown on right)
The alarm settings (upper/lower limit, ON/OFF) can be changed.
2
Use the
/
keys to change the displaying parameters.
3
The beds to be displayed can be filtered.
[This Display]: The beds monitored on the display unit currently displaying the
alarm settings for all beds will be displayed.
[Other Display]: The beds monitored on other display unit will be displayed.
[Nurse Team]: The beds registered for the selected nurse team will be displayed.
4
The parameters to be displayed can be filtered.
Select from [Basic]/ [Circ.]/ [Resp./Gas]/ [Arrhy.]/ [ST]/ [List]/ [Priority (Top)]/ [Priority (High)]/ [Priority
(Med.)]/ [Priority (Low)].
5
The specific parameters can be highlighted.
[Alarm ON]/[Alarm OFF]: The parameter which the alarm is set to ON, or OFF will be highlighted.
[Nurse Call ON]/[Nurse Call OFF]: The parameter which the nurse call is set to ON, or OFF will be
highlighted.
[OFF]: The highlight display will be cancelled.
6
The currently displayed data will be printed on the laser printer.
NOTE
If a laser printer is not connected, [Print] key will not function.
7
Use the
/
keys to change the displaying beds.
Alarm Occurrence
When the measurement data exceeds the alarm limit. or when arrhythmia is detected, or when connection failure of
the equipments occur, the alarm will be notified by message and sound.
When an alarm occurs, the numeric data will be displayed in reversed color or in 3D depending on the setting.
(Default: Reversed)
(
The waveform background will light in the color corresponding to the alarm priority. (Default: Lighting)
(
Maintenance Manual "Alarm" P7-2)
The alarm indicator will light.The flashing pattern can be changed.
(
Maintenance Manual "Alarm" P7-2)
The event key will be displayed on the home display for the alarm generated bed. By pressing the event key,
the alarm can be silenced for that bed.
When more than one alarms are generated, the alarm message with the higher priority will be displayed.
If more than one alarm with the same priority is generated, the message for the newest alarm will be displayed.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......