7-6
Chapter 7 Alarm Function
Alarm Limit Setup
Arrhythmia Alarm Setup
The arrhythmia alarm can be turned ON or OFF, and arrhythmia detection level can be set.
When all arrhythmia alarm is set OFF, <ARRHY OFF> message will be displayed.
Arrhythmia Alarm
NOTE
For the alarms such as "Asystole", "Run", "Pause", "Frequent", threshold levels needs to be
set.Other than ON/OFF of alarm, set also the threshold levels.
1
Press the [Menu], [Arrhy.] ("Alarm") key on the
Individual Bed Display.
The arrhythmia alarm setup screen will be
displayed.
2
Set the detection level.
Set using the dropdown list, numeric keys, or
displayed key selection.
3
Select ON/OFF for the alarm.
[ON]: Alarm will generate.
[OFF]: Alarm will not generate.
NOTE
Asystole, VF, VT alarm cannot be turned OFF.
HR Lower Limit for VT/Run/SVT
On the "Detail Setup" for arrhythmia alarm, HR lower limit for VT/Run/SVT can be set.
CAUTION
The settings for the "HR Lower Limit for VT", "HR Lower Limit for Run", "HR Lower Limit for
SVT" will be compared with the average HR of continuous VPC. Therefore, the displayed
HR value at alarm generation may be lower than the settings if it is just after the VT
detection, or if RUN with few continuous VPC is detected.
Arrhythmia Detection Level Setting
Item
Description
Item
Description
Asystole
3 sec. to 10 sec.
R on T
200 ms to 600 ms
Run
2 beats to 8 beats
SVT
2 beats to 10 beats
Pause
1.5 sec. to 5 sec.
Irregular RR
10, 15, 20%
Frequent
1 bpm to 50 bpm
S Frequent
1 bpm to 50 bpm
Ext Tachy
22 beats to 300 beats
Pacer Not Capture
80 ms to 480 ms
Ext Brady
20 beats to 295 beats
Pacer Not Pacing
20 bpm to 200 bpm
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......