6-13
Chapter 6 Admit/Discharge
Suspend Monitoring
Suspend Monitoring
This section explains the procedure to suspend and resume monitoring when a patient temporarily leaves the bed.
With this suspend monitoring function, data measurement, alarm generation, automatic measurement, and automatic
printing can be suspended without erasing any data and setup condition.
By using the monitor suspend label function, different labels in different colors according to the patient's destination
can be displayed during the monitoring suspended condition.
To remind the user to resume monitoring, alarm will generate after the preprogrammed duration (15 min./30 min./1
hr/1.5 hr/2 hr) for "Monitor Suspend Timer".
By selecting [ON] for "Auto Resume Monitoring", the monitoring will automatically start under the specified
condition. (
"Condition for Auto Resume Monitoring" P6-16)
CAUTION
During the monitoring suspended condition, the trend data and full disclosure waveform data
will not be acquired.
Resuming monitoring will also resume the suspended alarm.
Depending on the model type and software version of the bedside monitor, the monitor
suspend/resume operation will not synchronize between the bedside monitor and the central
monitor.
Suspending/resuming monitoring on the bedside monitor will not suspend/resume
monitoring on the central monitor.
Similarly, suspending/resuming monitoring on the central monitor will not suspend/
resume monitoring on the bedside monitor.
If monitoring is suspended on the bedside monitor, the data for that patient will not be
displayed on the central monitor. If monitoring is resumed on the bedside monitor, the
data for that patient will be displayed again on the central monitor.
The monitor suspend/resume operation on the bedside monitor can be synchronized to the
central monitor, but the monitor suspend operation on the central monitor cannot be
synchronized to the bedside monitor.
For details of the bedside monitor which is compatible to synchronizing the monitor suspend
operation, contact your nearest service representative.
When the monitoring is resumed, make sure that the monitoring is also resumed on the
central monitor.
REFERENCE
To display the detailed message during monitoring suspended condition, select ON for
"Monitor Suspend's Message Selection", and set the details under [Menu] > [Monitor
Suspend] ("Common Setup"). (
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......