![Fukuda CardiMax 2 Service Manual Download Page 75](http://html.mh-extra.com/html/fukuda/cardimax-2/cardimax-2_service-manual_2344835075.webp)
7-3
Chapter 7 Troubleshooting
Hardware Trouble
File
Hardware Trouble
DANGER
While operating the equipment, do not unplug the AC power cable or remove the battery. To
cease the operation, use the power supply switch.
It may cause irreparable damage to the system file.
CAUTION
When replacing parts, it is recommended to replace in board unit. The electric parts soldered
on to the board are very small surface-mounted components, and may damage the board
or other parts when replacing.
If the power cannot be turned OFF using the power ON/OFF key, press down the key for
about 10 seconds to turn OFF the power forcibly.
Always remove the batteries before checking the board and connectors. Otherwise, it may
cause damage to the equipment.
Message
Cause
Solution
<Failed to format (Media).>
Formatting of the media has failed.
The failure of the media can be considered.
Contact your local Fukuda Denshi service
representative.
<There is not enough space on
(media). Please replace the media
with another one.>
There is not enough space on the media to
save the data.
Replace with a media with enough capacity.
<Cannot write due to insufficient
space on (media).>
The data could not be saved on the media as
the maximum number of recordable data has
been reached.
When this message is displayed, the data is
not saved.Insert a new media and save the
data again.
<Cannot store data.>
Failed to save the data to the media.
The failure of the media can be considered.
Write down the error code at the end of the
message and contact your local Fukuda
Denshi service representative.
<Please set (media).>
The media is not set.
Set the media.
<Failed to read from the file.>
Failed to read the data from the file.
The failure of the media can be considered.
Write down the error code at the end of the
message and contact your local Fukuda
Denshi service representative.
<The examination data was not
saved.>
The examination data was not saved when
proceeding to the next examination.
Save the examination data before proceeding
to the next examination.
<Failed to delete data.>
Failed to delete ECG data.
The failure of the media can be considered.
Contact your local Fukuda Denshi service
representative.
<No data selected.>
A process was performed without selecting
data.
Select data before performing the process.
<Failed to save the settings.>
Failed to save the settings to the media.
The failure of the media can be considered.
Contact your local Fukuda Denshi service
representative.
<Failed to load settings.>
Failed to load the settings from the media.
Make sure to use the correct media to which
settings are saved.
If failure occurs even though the correct media
is used, it may be defective. Contact your local
Fukuda Denshi service representative.
bl5667
bl5958
Summary of Contents for CardiMax 2
Page 4: ...2 ...
Page 10: ...Contents ...
Page 26: ...Chapter 1 General Description Contents ...
Page 34: ...Chapter 2 Operational Description Contents ...
Page 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Page 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Page 48: ...Chapter 4 Spare Parts List Contents ...
Page 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Page 68: ...Chapter 6 Software Update Contents ...
Page 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Page 72: ...Chapter 7 Troubleshooting Contents ...
Page 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Page 82: ...Chapter 8 Maintenance Contents ...
Page 104: ...Chapter 9 Maintenance and Inspection Contents ...
Page 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Page 116: ...Chapter 10 Appendix Contents ...