Safety
Design Specifications
This equipment meets the requirements of IEC 60601-1(2005) +Am1(2012) "Medical electrical equipment – Part 1:
General requirements for basic safety and essential performance" and is classified as follows.
1 Protection against electrical shock
Class I, Internally Powered Device
2 Degree of protection against electrical shock
Type CF Applied Part
3 Degree of protection against ingress of water
No special protection against ingress of water (IPX0) .
4 Classification according to the degree of safety when using in the presence of air or flammable anesthetic gas
or oxygen or nitrous oxide and flammable anesthetic gas.
To be used in environments without flammable anesthetic gas or other flammable substances
5 Operation Mode
Can be used continuously
6 Classification of Sterilization Method
Not intended to be sterilized
7 Target
Neonate, Child, Adult
Preventive Maintenance
The purpose of preventive maintenance is to ensure that the equipment is in safe operating conditions at all times and
prevent potential future problems.
Perform preventive maintenance at least once every three months.
Inspect the main unit and all connectors and cables for any external damage.
Perform maintenance if one of the following should occur.Do not repair or perform maintenance when the equipment
is in use.
When the equipment gets a strong impact due to dropping, etc.
When liquid spills on or into the equipment
When the equipment does not function normally
When the enclosure of the equipment is cracked, broken or damaged
Cables such as the power cable and lead wires have deteriorated in performance
Check all cables, equipments and accessories for damage, earth impedance, leakage current and accuracy.
Precautions when Using with Other Equipment
Non-Explosion Proof
Never operate the equipment where flammable gas or fluid such as anesthetic, oxygen, and hydrogen are used.
Explosion or fire may result.
Never operate the equipment in the presence of flammable anesthetics, high concentration of oxygen.
This may cause an explosion or fire.
Never use the equipment in the hyperbaric oxygen therapy chamber.
This may cause an explosion or fire.
bl5637
bl5638
DANGER
Summary of Contents for CardiMax 2
Page 4: ...2 ...
Page 10: ...Contents ...
Page 26: ...Chapter 1 General Description Contents ...
Page 34: ...Chapter 2 Operational Description Contents ...
Page 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Page 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Page 48: ...Chapter 4 Spare Parts List Contents ...
Page 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Page 68: ...Chapter 6 Software Update Contents ...
Page 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Page 72: ...Chapter 7 Troubleshooting Contents ...
Page 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Page 82: ...Chapter 8 Maintenance Contents ...
Page 104: ...Chapter 9 Maintenance and Inspection Contents ...
Page 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Page 116: ...Chapter 10 Appendix Contents ...