9-5
Chapter 9 Maintenance and Inspection
Electrical Safety Inspection Procedure
Other
Electrical Safety Inspection Procedure
Electrical safety-related testing methods and measurement equipment are stipulated in safety test standards (EN
60601-1). A dedicated safety test measuring instrument is required for performing this measurement . Check that the
measurement results do not exceed the threshold value and record the numerical values.
Safety Evaluation Test Device
Testing and inspection items in this manual are described based on the assumption that a dedicated electrical safety
tester with its accuracy controlled is used. Considering ease of test and ensuring accuracy, it is recommended to use
a dedicated safety test measuring instrument.
Contact our service personnel for more information on safety test measuring instrument.
Check Items
Electrical safety check items include the following types.
Measurement is performed under the following two conditions.
Item
Inspection Item
Procedure
Criteria
Other
1) Power supply
Check whether the device can be turned
on/off both with AC power and battery
power.
The device should be turned on/off both
with AC power and battery power.
2) Power cable
Perform continuity test using a tester.
Perform a continuity test using a tester
and check that the equipment is
conducting.
(1)
Earth Leakage Current
Leakage current that runs to the protective earth conductor.
(2)
Touch Current
Leakage current that runs through the ground from the enclosure of the device
(3)
Patient Leakage Current
(From patient connection to earth)
Leakage current that runs through the ground via patient from the attaching part
There are two types of measuring method - direct current and alternating current.
(4)
Patient Leakage Current
(Current when external voltage is
applied to SIP/SOP)
Leakage current that runs through the ground via patient from the attaching part by
means of the supply voltage carried on the signal input/output
There are two types of measuring method - direct current and alternating current.
(5)
Patient Leakage Current
(When external voltage is applied to
patient connection of Type F Applied
Part)
Leakage current that runs through the ground from the device by means of the supply
voltage carried on the F type attaching part through patient
(6)
Patient Auxiliary Current
Current that runs between attaching parts through patient
There are two types of measuring method - direct current and alternating current.
bl5867
(1)
Normal Condition
Measurement in normal condition
(2)
Single Failure Condition
Condition under which a single fault occurs
Summary of Contents for CardiMax 2
Page 4: ...2 ...
Page 10: ...Contents ...
Page 26: ...Chapter 1 General Description Contents ...
Page 34: ...Chapter 2 Operational Description Contents ...
Page 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Page 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Page 48: ...Chapter 4 Spare Parts List Contents ...
Page 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Page 68: ...Chapter 6 Software Update Contents ...
Page 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Page 72: ...Chapter 7 Troubleshooting Contents ...
Page 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Page 82: ...Chapter 8 Maintenance Contents ...
Page 104: ...Chapter 9 Maintenance and Inspection Contents ...
Page 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Page 116: ...Chapter 10 Appendix Contents ...