6. SpO
2
Monitoring (LX-7230KM)
35
SpO
2
Monitoring (LX-7230KM)
LX-7230KM, which has a built-in Konica Minolta
®
SpO
2
module, is described in
this section.
Regarding the LX-7230N, which has a built-in Nellcor
®
SpO
2
module, refer to
page 18.
< SpO
2
probes for LX-7230KM >
The optional SpO
2
probes available for LX-7230KM are as follows.
The following table shows applicable patient and proper site for each SpO
2
probe. Select the proper one depending on the purpose and intended use.
Sensor Types
Applicable Patient
Applied Site
Finger Clip
SR-5C
Adult and Pediatric
(weight of 30Kg and over)
Finger
Spot Check
SP-5C
Adult and Pediatric
(weight of 30Kg and over)
Finger
Universal Probe
UD-5C
Adult and Pediatric
(weight of 40Kg and over)
Finger
Adult and Pediatric
(weight of 15Kg and over)
Finger
Personal Probe
SD-5C(25),
SD-5C(20)
Pediatric and Infant
(weight of 5 to 15Kg)
Big toe or the root
of the big toe
WARNING
For SpO
2
monitoring, always use the probe/relay cable specified by
Fukuda Denshi. If any other probe/relay cable is used, high temperature
rise of the sensor may place the patient in danger of burns in the worst
case.
As with all medical equipment, carefully route cables to reduce the
possibility of patient entanglement and strangulation.
Summary of Contents for LX-7230
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Page 76: ...7 Measurement 56 Blank Page ...
Page 86: ...9 Other Setting Items 66 Blank Page ...
Page 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Page 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Page 111: ......