4-7
Chapter 4 Connection to the External Devices
Setup for the External Device Connection
Alarm Output Setup
The alarm can be output from the Status II connector on the AC Unit (DSA-81).
1
Press the [Menu], [Setup], [Initial Settings], [External Device], [AC Unit], [Status Output] keys.
The "Status Output" setup menu will be displayed.
2
Set the "Alarm Level".
[OFF]: Alarm will not be output.
[APNEA]: Apnea alarm will be output.
[Level H]: Level H alarm will be output.
[Level H,M]: Level H, M alarm will be output.
[Level H,M,L]: Level H, M, L alarm will be output.
3
Set the "Output Logic".
[Positive Logic]: The positive logic alarm will be output.
[Negative Logic]: The negative logic alarm will be output.
[Pulse]: A square wave of 440 ms cycle will be output.
NOTE
"Status I/O Signal (Status II Connector)" P6-13 for connector pin
assignments of the alarm output.
The equipment status alarm will be output as level L.To output the equipment status
alarm, select [Level H,M,L].
On the DS-8007 system, the status output settings are saved on the AC Unit (DSA-81).
Analog Output Setup
On the DS-8007 system, analog output of ECG waveform and BP waveform is possible. For "Analog Output 3",
synchronized signal (HR or RR) can be also output.
The BP waveform for analog output can be selected from the measured waveforms on the DS-8007 multiamplifier
connector.
On the "Analog Output" setup menu, initial settings for display/printing can be performed.
1
Press the [Menu], [Setup], [Initial Settings], [External Device], [Analog Output] keys.
The "Analog Output" setup menu will be
displayed.
2
Set the "Analog Synchronized Signal Output".
Select from [ON]/[OFF].
3
Set the "Analog Output 1", "Analog Output 2".
Select from [Selected ECG Lead]/[Displayed ECG
Lead]/[Multiparameter Connector 1-1]/
[Multiparameter Connector 1-2]/[Multiparameter
Connector 2-1]/[Multiparameter Connector 2-2].
2
3
2
3
4
4
5
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......