Chapter 3 Using the External Media
Formatting the Full Disclosure Waveform Card
The following items will not be backed up/copied.
Setup Data
Time/Date
Telemeter Setup
(The settings are saved in the connected telemetry transmitter module.)
Patient Data
OCRG Data
CO Measurement Result
Formatting the Full Disclosure Waveform Card
In this section, formatting of SD card to be used for storing the full disclosure waveform is explained.
By formatting the specified card for full disclosure waveform, the card can be used for storing the full disclosure
waveform data. By inserting the formatted card to the card slot, storing of the full disclosure waveform data will
automatically start, and reviewing of the full disclosure waveform will become possible.
CAUTION
The full disclosure waveform card formatted on other bedside monitors and central monitors
cannot be used on this equipment.
The full disclosure waveform card formatted on this equipment cannot be used on other
bedside monitors and central monitors.
While saving the full disclosure waveform to the card, avoid removing/inserting the card
beyond necessity.
It will take about 3 minutes to format the full disclosure waveform card. Do not format the
card during monitoring as all operation will not be possible during the format process.
During the format process, do not turn OFF the power, or enter into standby condition, or
remove the card. It may damage the card.
Setup Data
Basic Setup
Stores the current setup.
Alarm
Stores the alarm ON/OFF and alarm limit settings.
Parameter Setup
Stores the monitoring condition (size, lead, etc.) for the parameter.
Data Review/Waveform Review/
Calculation
Stores the settings for each review data.
Initial Settings
Stores the current setup.
Setup Data
Data
Description
Patient Data
Data
Description
Patient Information
Stores the patient information such as name, ID, age, sex, pacemaker usage, patient
classification.
Graphic Trend Data
Stores 24 hours of graphic trend data.
Tabular Trend Data
Stores 24 hours of tabular trend data.
Recall
Stores 300 recall data.(10 data if SD card is not used.)
Hemodynamic Data
Stores 10 measurement data.(5 data if SD card is not used.)
Lung Function Data
Stores 256 measurement data.(5 data if SD card is not used.)
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......