Chapter 2 Network System Construction
Wired Network System
Chapter 2 Network System Construction
Wired Network System
In this section, connection and setup procedure for wired network is explained.
A wired network system can be constructed by using the LAN cable. Maximum of 48 beds for the DS-LAN II
network, maximum of 100 beds for the DS-LAN III network can be connected. The central monitor corresponded to
each wired network is required and the central monitor with the central ID "1" will function as the network
administrator.
DS-LAN Connection
WARNING
Do not connect unspecified equipment to the wired network.
Do not mix devices with DS-LAN II and DS-LAN III setting in the same wired network.The
network may cease and proper monitoring may not be possible.
CAUTION
On a wired network, the alarm generated on the bedside monitor will be output to the
network with a maximum delay of 1 second, and to the central monitor with a total delay of
2.5 seconds.
When connecting to the DS-LAN network, perform "DS-LAN Setup" under [Initial
Settings>System>DS-LAN] and restart the system before connecting the LAN cable.
Use a repeater HUB for DS-LAN II network and a switching HUB for DS-LAN III network.
On the DS-8007 system, the Room ID/Bed ID are saved on the AC Unit (DSA-81).
On the AC Unit (DSA-81), the default setting of Bed ID is "000". If connected to a wired
network with the bed ID unchanged, monitoring on the central monitor will not be possible.
When connecting to a wired network, make sure that there are no other bedside monitors
with the same ID. If there is more than one bedside monitor with the same bed ID, the
duplicated bedside monitors cannot be monitored on the central monitor.
When connecting to the DS-LAN II network, set the ID in the range from 001 to 048. When
connecting to the DS-LAN III network, set the ID in the range from 001 to 100.
If the measurement unit of CO
2
concentration is mmHg and [99 mmHg] is selected for
"Upper Limit of CO
2
(mmHg) Transmission" under [Initial Settings>System>DS-LAN or
Telemeter], the CO
2
value of 100 mmHg or above will be transmitted as 99 mmHg.
Configure the network as specified by Fukuda Denshi. If the equipment of different network
type is connected, malfunction may occur to the whole network system.
By connecting a Ethernet branch cable to the DS-LAN connector on the AC Unit (DSA-81), a wired network system
can be constructed.
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......