Chapter 1 Installation of the Unit
Precautions for Installing the Equipment
Chapter 1 Installation of the Unit
Precautions for Installing the Equipment
This section describes the environmental condition to use this equipment.
CAUTION
The installation of this equipment should be performed by our service representative or a
person who is well acquainted with this equipment.
Avoid stacking multiple equipments or placing anything on the equipment during operation.
Install this equipment in a place where power supply cable can be easily disconnected.
If any abnormality is found on the equipment, immediately turn OFF the power, and
disconnect the power supply cable from the outlet.
Do not soak or immerse the equipment in liquids.
Operating Environment
The following environmental conditions should be observed when operating the equipment.
Surrounding Temperature: 10
°
C to 40
°
C
Relative Humidity: 30% to 85% (non-condensing)
Atmospheric Pressure : 70 kPa to 106 kPa
This equipment is intended for patient monitoring in NICU, ICU, CCU, surgery, emergency room and ward.
Do not use in MRI environment or in a home-care setting.
The power source should fulfill the following condition.
Use a hospital grade outlet (3-pin grounded outlet).
Verify power voltage and frequency before connecting to an AC power source.
Use the power source that can provide adequate power to the device.
Refer
Operation Manual "Main Unit: DS-8007" P14-1 for power voltage, frequency, and power
consumption.
Pay attention to install or store the equipment in proper location. Do not install or store in the following
locations.
where chemicals are stored or gas may generate
where the equipment will be subject to splashing water or humidity from a nebulizer or vaporizer
where the equipment will be subject to direct sunlight
where the equipment will be subject to inclination, vibration, or shock.
Ensure proper ventilation to cool the device.
A minimum space of 5 cm is required between the rear side of the monitor and the wall.If the monitor is
embedded in a wall or surrounded by a wall, a minimum space of 10 cm is required.
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......