Precautions about the Ventilator Monitoring
Precautions about the Ventilator Monitoring
The ventilator alarm sound is set to OFF at factory default setting.
The alarm sound can be turned ON on the Tone/Volume setup screen.
If the DS-8007 System does not generate an alarm even though the ventilator is generating an alarm, or if any
other malfunction occurs, immediately check the ventilator, this equipment, cable, and replace the cable if
necessary. If the malfunction persists, stop using the equipment.
The alarm generation on the DS-8007 System is not guaranteed if the alarm other than the specified one
generates at the ventilator.
(
Maintenance Manual "Ventilator Alarm Input" P4-1)
The ventilator operation should be performed by well-trained and authorized personnel.
When connecting this equipment and a ventilator, use only the specified connection cable.
Verify that this equipment and the ventilator are properly connected.
When connecting the cable, verify that the main power of this equipment and the ventilator are OFF.
During ventilator monitoring, if the DSA-81 is disconnected from the DS-8007, the ventilator alarm will not
generate.
Precautions about the SpO
2
Sensor
Danger of Burn Injury Caused by the SpO
2
Sensor
When monitoring SpO
2
, make sure to use only the specified sensor/relay cable. If any other sensor/relay cable
is used, a high temperature rise of the sensor may place the patient in danger of burns.
If there are any questions regarding the sensor/relay cable use for SpO
2
measurements of this equipment, please
contact Fukuda Denshi service representative.
Precautions about the NIBP Cuff
Some of the NIBP cuffs used for this equipment contain natural rubber latex which may cause allergic
reactions.
(FDA: Medical Alert on Latex Products, "Allergic Reactions to Latex-Containing Medical Devices", Food &
Drug Administration, 10903 New Hampshire Avenue, Silver Spring, MD 20993, 1991.)
Precautions about Disposing of the Equipment, Accessories, or Components
When disposing of this equipment, accessories, or components, use an industrial waste distributor. Do not
dispose of as ordinary waste.
When disposing of the battery, separate it from other wastes and contact your nearest service representative.
WARNING
CAUTION
DANGER
CAUTION
CAUTION
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......