xviii
Safety
Precautions for Using the Equipment
Depending on the central monitor model type, the ST display will be distorted if the ECG lead (ECG1 or ECG
2) is changed on the DS-8007 System system. Redrawing the ST display will return the display to normal.
On the central monitor, the respiration waveform and RR value based on the RR/APNEA alarm source selected
on the DS-8007 System system will be displayed. The RR and APNEA monitored on the central monitor and
the DS-8007 System will be the same.
Wireless Network System
When monitoring a patient using wireless telemetry, make sure the patient data is properly received at the
central monitor. Pay special attention when the channel ID at the bedside monitor is changed.
A password can be set to access the channel ID setup menu to allow only the telemetry channel administrator
to change the channel ID.
Some type of wireless combinations may generate interference with other telemetry.
Before selecting a channel, verify it will not interfere with other channels.
Inform the supervisor of the use of telemetry channels to avoid interference with other telemetry.
If transmitters are used in a neighboring medical facility, your facility and the neighboring facility must make
agreements on the setting of the telemetry channels to prevent telemetry interference.
Precautions about the Telemetry
When performing telemetry transmission, configure the display so that the numeric data corresponded to the
waveform is displayed. If not, the displayed waveform or numeric data may not be transmitted.
The setup of channel ID and group ID should be performed only by the telemetry channel administrator or our
service representative. Users should not perform this procedure as malfunction may occur.
When the measurement unit of BP is "kPa", corresponding waveform and numeric data will not be transmitted.
When using a wireless network, use "mmHg" for the BP measurement unit.
BP waveform with a scale above the set scale can not be properly transmitted. When transmitting the BP
waveform, pay attention to the displayed BP scale.
If the measurement unit of CO
2
concentration is mmHg and [99mmHg] is selected for "CO
2
(mmHg) Upper
Limit of Transmission" ([Initial Settings]>[System]>[Telemeter]), the CO
2
value of 100mmHg or above will
be transmitted as 99mmHg.
RTC and Data Backup
This equipment is equipped with a built-in clock.When the power of this equipment is turned OFF, this clock
is backed up by a lithium primary battery.If incorrect time is displayed when turning ON the power, a low
battery may be the cause. In such case, contact Fukuda Denshi service representative for replacing the battery.
DANGER
WARNING
CAUTION
CAUTION
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......