xvii
Safety
Precautions for Using the Equipment
Wired Network (DS-LANII/ DS-LANIII)
Do not connect unspecified device to the wired network.
Do not mix devices with DS-LANII and DS-LANIII setting in the same wired network. The network may cease
and proper monitoring may not be possible.
If performing wired network transmission, configure the display so that the numeric data corresponded to the
waveform is displayed. If not, the displayed waveform or numeric data may not be transmitted.
On the DS-8007 system, the following settings are saved on the AC Unit (DSA-81).
Room ID, Bed ID, DS-LAN II/III Setup, Pat. ID Transmission Start Position, Hemodynamic Data
Synchronization, CO
2
(mmHg) Upper Limit of Transmission
On the AC Unit (DSA-81), the default setting of Bed ID is "000". If connected to a wired network with the bed
ID unchanged, monitoring on the central monitor will not be possible.
When connecting to a wired network, make sure that there are no other bedside monitors with the same ID. If
there is more than one bedside monitor with the same bed ID, the duplicated bedside monitors cannot be
monitored on the central monitor.
When connected to the DS-LAN II network, set the Bed ID in the range from "001" to "048" .
When connected to the DS-LAN III network, set the Bed ID in the range from "001" to "100" .
There are following restrictions when connecting the DS-8007 System to the wired network.
The BP measurement unit setting should be the same for all central monitors and bedside monitors. If the
setting is different among the monitors, data such as BP waveform, BP numeric data, NIBP numeric data,
NIBP list will not be transmitted.It will be treated as not measured data, and will not be displayed on the
central monitor.The alarm limit setup from the central monitor cannot be performed either.
On the DS-LAN II network, the arrhythmia alarm of Tachy, Brady, Couplet, Pause, Trigeminy, ExtTachy,
ExtBrady, RR IREG, Prolong RR, Triplet, Multiform, VENT Rhythm, Not Capt, Not Pacing, S Couplet,
SVT, SVPC, S Frequent will not be transmitted.
On the DS-LAN II network, arrhythmia alarm of "SLOW VT" will be transmitted as "VT" .
On the wired network, measurement data and alarm of TEMP3 to 6 will not be transmitted. Also, the
displayable waveform, numeric data, alarm differs depending on the connected central monitor.Refer also to
the operation manual for the respective central monitor.
The PR_IBP alarm will not be transmitted to the central monitor.
If the "RR/APNEA alarm source" is other than [Impedance] (or, if [Auto] selects a setting other than
[Impedance]), the RESP waveform will not be transmitted on a wired network.
If the "RR/APNEA Alarm Source" setting is other than [CO
2
] (Or, if [Auto] selects a setting other than
[CO
2
]), the CO
2
waveform will not be transmitted on a wired network.
For the numeric data displayed as "xxx", maximum or minimum value of measurable range will be
transmitted.
The numeric data displayed as "--- " will be treated as not measured data.
If the measurement unit of CO
2
concentration is mmHg and [99mmHg] is selected for "CO
2
(mmHg) Upper
Limit of Transmission" ([Initial Settings]>[System]>[DS-LAN]), the CO
2
value of 100mmHg or above will
be transmitted as 99mmHg.
As the DS-8007 System do not have the arrhythmia template display and 12-lead ST display function,
waveforms and other data will not be displayed for these displays on the central monitor connected to the DS-
LAN network.
When connected to the wired network, the time/date will synchronize with the central monitor.Even if the time/
date is changed on the DS-8007 System, it will be corrected to the time/date of the central monitor.
WARNING
CAUTION
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......