vi
Safety
Precautions about the Network System
Precautions about the Network System
Medical Telemetry
Precautions about the Installation
The medical institution (hereinafter referred as "Institution" ) must decide the telemetry installation plan for
the medical institution in order to prevent interference between transmitters (telemetry based on destination
country's radio law). When telemetry has already been installed and been used, radio format, frequency, and
antenna power are required to be examined to prevent interference.
When using telemetry which requires zone location, the institution is to set up the zones as an operation unit
for each transmitter to prevent electronic interference between telemetry throughout the Institution.
When using telemetry which requires zone location, display and identify each prepared zone in the equipment.
When laying receiver antenna for each transmitter, the Institution has to examine the installation so that
electronic interference does not occur.
Based on the above examination result, the Institution should place each receiver antenna as required.
Precautions about the Management
The institution appoints a person to manage the wireless channels for the whole medical institution. And when
using telemetry which requires zone location, the Institution should nominate a person to manage the wireless
channels in each zone (a "Zone Manager" ). However, when using such telemetry in a local medical institution,
one person can perform both functions.
Select a telemetry manager who understands the characteristics and functionality of telemetry systems, and is
skilled in operating telemetry.
When installing telemetry, the Overall Manager and the Zone Manager have to understand the precautions for
use of the telemetry in advance.
The Overall Manager takes responsibility of wireless channel management and transmitter storage for the
whole Institution by giving proper instruction.
The Overall Manager should create a management log (hereinafter referred to as the "log" ), which contains a
list of the management status of the wireless channels for the whole Institution. When changing a wireless
channel, register it in the log and give proper instructions to the Zone Manager or to the user.
The Zone Manager assumes responsibility for managing the wireless channels, storing, and managing
telemetry.
The Zone Manager assigns the transmitter to the user, and provides enough education for use inside the zone.
The telemetry user verifies operation of the transmitter/receiver before use.
The telemetry user, if using the telemetry in a zone location, follows the instructions of the Zone Manager for
the zone and gives instructions to the patient if required.
When interference or breakdown occurs in telemetry communication, the user is required to inform the zone
manager and the overall manager of the problems. The Zone Manager and Overall Manager are to deal with
the problem properly and/or contact their nearest Fukuda Denshi representative for service.
CAUTION
CAUTION
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......