Electromagnetic Compatibility
The performance of this device under electromagnetic environment complies with IEC60601-1-2
(2001).
Precautions for Safe Operation under Electromagnetic Influence
C A U T I O N
If any sorts of electromagnetic wave, magnetic field, or static electricity exist
around the device, noise interference or malfunction of the device may occur.
If any unintended malfunction or noise occurs during monitoring, check the
magnetic influence and take appropriate countermeasures.
The following are examples of the common cause and countermeasures.
Cellular Phone
The radio wave may cause malfunction to the device.
Cellular phones and radio sets should be turned off in the room (building)
where medical device is located.
Static Electricity
In a dry environment (room), static electricity is likely to occur. Take the
following countermeasures.
Both operator and patient should remove any static electricity before
entering the room.
Humidify the room.
Lightning
A lightning nearby may induce excessive voltage to the equipment.
If any danger is suspected, use the uninterruptible power supply
system.
High frequency noise interference from other device through the power
outlet
Check where the noise is originated and remove it using filtering
device, etc.
Stop using the device that is originating the noise.
Use other power outlet.
EMC Guidance
This equipment complies with IEC60601-1-2 (2001). However, if portable transmitter or wireless
LAN equipment is used extremely nearby, the electromagnetic influence may largely exceed the
compliance level and may cause unexpected phenomenon such as noise interference on the
waveform, etc.
Therefore, this equipment should be used in a location specified by each medical institution.
If any unexpected noise interference on the waveform or failure to the peripheral device occurs,
stop using the equipment and follow the instruction of the technician.
The following is the information relating to EMC (Electromagnetic Compatibility).
(When using this equipment, verify that it is used within the environment specified below.)
xxvii
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...