Replacing the Air Filter
The cooling fan air filter on the DS-7000 and Multigas Unit is a consumable product.
Continuous operation of the fan will cause the air filter to inhale unclean air inside the equipment,
reduce the cooling effect, and may damage the inner parts.
Clean the air filter, or replace it with a new air filter every 3 months.
W A R N I N G
After washing the air filter with neutral detergent, dry it completely before
reattaching. If moisture remains on the air filter, it may damage the
equipment.
The air filter must be attached after cleaning / replacing. If the equipment
is used without it, it may become un-functional.
●
Cleaning and Replacing the Air Filter on the DS-7000 Main Unit
1
Turn the cooling fan cover to the “UNLOCK” direction, and take it off when the
▲
mark
matches the arrow mark on the monitor.
2
Take out the air filter and clean it, or replace it as necessary.
To clean the air filter, shake the dust off, or wash it off with neutral detergent.
Air Filter
3
Reattach the cooling fan cover.
Turn to the “LOCK” direction and attach it so that the
▲
mark and arrow mark aligns.
7
-
22
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...