Ex-AGT Alarm
The upper/lower limit can be set in 0.1% increments.
Lower
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 0.0 to 4.5%
DES: 0.0 to 17.5%
SEV: 0.0 to 7.5%
Turns OFF if the value below the adjustable range is
set.
Upper
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 0.5 to 5.0%
DES: 0.5 to 18.0%
SEV: 0.5 to 8.0%
Turns OFF if the value above the adjustable range is
set.
Auto
Automatically sets the upper limit to +0.8%, and lower
limit to –0.8% to the current value.
●
MAC Value Alarm
The upper limit can be set in 0.1 increments.
Lower
Limit
(No setting)
Upper
Limit
Range: 0.1 to 9.9.
Turns OFF if 9.9 or above is set.
To maintain the alarm setting even after the power is turned OFF or after the discharge procedure,
store the setting to one of the alarm modes, or select Backup for “Alarm” on the “Backup at
Discharge” menu (Monitor Setup).
Reference
For the alarm mode setup procedure, refer to “8. System Configuration Alarm Mode”.