6
-
87
●
N
2
O (Nitrous Oxide) Alarm
In-N
2
O Alarm
The upper/lower limit can be set in 2% increments.
Lower
Limit
Range: 0 to 98%.
Turns OFF if 0% or below is set.
Upper
Limit
Range: 2 to 100%.
Turns OFF if 100% or above is set.
Auto
Automatically sets the upper limit to +10%, and lower
limit to –10% from the current value.
Ex-N
2
O Alarm
The upper/lower limit can be set in 2% increments.
Lower
Limit
Range: 0 to 98%.
Turns OFF if 0% or below is set.
Upper
Limit
Range: 2 to 100%.
Turns OFF if 100% or above is set.
Auto
Automatically sets the upper limit to +10%, and lower
limit to –10% from the current value.
●
Anesthetic Gas Alarm
In-AGT Alarm
The upper/lower limit can be set in 0.1% increments.
Lower
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 0.0 to 4.5%
DES: 0.0 to 17.5%
SEV: 0.0 to 7.5%
Turns OFF if the value below the adjustable range is
set.
Upper
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 0.5 to 5.0%
DES: 0.5 to 18.0%
SEV: 0.5 to 8.0%
Turns OFF if the value above the adjustable range is
set.
Auto
Automatically sets the upper limit to +0.8%, and lower
limit to –0.8% to the current value.
6
Multigas
Data (PHASEIN
®
IRMA Probe)
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...