NIBP Measurement at Alarm Occurence
The NIBP measurement can be started at alarm generation of specified parameter.
1
Press the NIBP Measurement at Alarm Occurrence on the NIBP display.
ON will start NIBP measurement at alarm generation of
the specified parameter.
When ON is set, select the parameters to start the NIBP
measurement. More than one parameter can be selected.
The NIBP measurement at alarm occurrence will be performed on the condition that at least one NIBP
measurement is performed since the power is turned ON, and 10 seconds or more has elapsed since
the last NIBP measurement.
NIBP Monitoring Condition Setup
Pressing the Setup key on the NIBP configuration menu allows setting the NIBP monitoring
condition.
●
Patient Classification Selection
The patient classification affects the NIBP measurement accuracy.
It also affects the accuracy of HR and RR measurement and the delay time to generate the alarm.
Adult
Child
Neonate
NIBP Measurement
Range
10 to 280mmHg
10 to 180mmHg
10 to 130mmHg
HR
0bpm, 12 to 300bpm
0bpm, 30 to 300bpm
Monitor
0.5 to 40Hz
1.6 to 40Hz
ESIS
1.6 to 15Hz
1.6 to 15Hz
Filter
Mode
Diagnosis
0.05 to 100Hz
1.6 to 100Hz
Impedance Respiration
1.5Hz
2.5Hz
Alarm Delay
5sec.
0sec.
The alarm delay duration is provided to prevent the alarm to generate frequently. The alarm will not
generate during the alarm delay duration even if the alarm threshold is exceeded.
The parameters which alarm generation will be delayed are HR/PR, BP, RR, SpO
2
, TEMP, TACHY,
and BRADY.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...