Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the DSL-8001 System
The DSL-8001 system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The electromagnetic interference can be prevented by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and
the DSL-8001 system as recommended below, according to the maximum output power of the
communications equipment.
Separation Distance according to Frequency of Transmitter (m)
Rated Maximum
Output Power of
Transmitter(W)
150kHz to 80MHz
d = 1.2
P
80MHz to 800MHz
d = 1.2
P
800MHz to 2.5GHz
d = 2.3
P
0.01 0.12 0.12
0.23
0.1 0.38 0.38
0.73
1 1.2
1.2 2.3
10 3.8 3.8
7.3
100 12 12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance in meters (m) can be determined using the equation applicable to the frequency of the
transmitter. In the equation, P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1
: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
Note 2
: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Summary of Contents for DSL-8001
Page 4: ...Blank Page ...
Page 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 44: ...2 6 Blank Page ...
Page 60: ...3 16 Blank Page ...
Page 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Page 64: ...4 4 Blank Page ...
Page 72: ...6 6 Blank Page ...
Page 102: ...7 30 Blank Page ...
Page 148: ...11 6 Blank Page ...